UV-1100 SPECTROPHOTOMETER
UV-1100 spectrophotometer is the ideal instrument for education and QC laboratories.Using your standard sample solutions,you can get a standard curve on the large LCD screen.They are widely used in colleges and enterprises for general quantitative analysis and experiments.
Features
1.Large LCD screen (128*64 Dots).
2.Can display total 50 groups of data,3 groups per screen.Can display standard curve and the curve equation.
3.System can also save the test results,total 200 groups of data and 200 standard curves can be saved,it is convenient for check and reload.
4.Data can be restored after a sudden power cut.
5.Auto setting wavelength.
6.Tungsten lamp & Deuterium lamp can be turned on/off individually to extend lifetime.
7.Pre-aligned design makes it convenient to change lamps.
8.Large sample compartment,it can accommodate 5-100mm path length cuvettes with optional holders,a variety of optional accessories are available.
9. The free application software M.Wave Basic provides complete control of the spectrophotometer through the Built-in USB port, you can achieve the following functions:
Ⅰ.Quantitative Ⅱ.Kinetics Ⅲ.Photometric Test
Functions:
1.Basic Mode:Absorbance,Transmittance,Concentration
2.Quantitative
Specifications:
Model | UV-1100 |
Wavelength Range | 200-1000nm |
Spectral Bandwidth | 4nm |
Optical System | Single Beam,Grating 1200 lines/mm |
Wavelength Accuracy | ±2nm |
Wavelength Repeatability | 1nm |
Photometric Accuray | ±0.5%T or ±0.004A@1A |
Photometric Range | 0-200%T,-0.3-3A,0-9999Conc |
Stray Light | 0.2%T |
Stability | ±0.002A/h@500nm |
Detector | Silicon Photodiode |
Standard Cell Holder | 4-position 10mm cell changer |
Sample Compartment | Standard 10mm pathlength cuvette |
Light Source | Tungsten & Deuterium lamp |
Output | USB Port & Parallel Port (Printer) |
Power Requirement | AC 110/220V 50/60HZ |
Dimension(L*W*H) | 440*387*180mm |
Weight | 14kg |
V-1200 SPECTROPHOTOMETER
V-1200 spectrophotometer is the ideal instrument for education and QC laboratories.Using your standard sample solutions,you can get a standard curve on the large LCD screen.They are widely used in colleges and enterprises for general quantitative analysis and experiments.
Features
1.Large LCD screen (128*64 Dots).
2.Can display total 50 groups of data,3 groups per screen.Can display standard curve and the curve equation.
3.System can also save the test results,total 200 groups of datas and 200 standard curves can be saved,it is convenient for check and reload.
4.Data can be restored after a sudden power cut.
5.Auto setting wavelength.
6.Tungsten lamp can be turned on/off individually to extend lifetime.
7.Large sample compartment,it can accommodate 5-100mm path length cuvettes with optional holders,a variety of optional accessories are available.
8. The free application software M.Wave Basic provides complete control of the spectrophotometer through the Built-in USB port,you can achieve the following functions:
Ⅰ.Quantitative Ⅱ.Kinetics Ⅲ.Photometric Test
Functions:
1.Basic Mode:Absorbance,Transmittance,Concentration
2.Quantitative
2.1 Standard Curve
Maximum of 9 standard samples can be used to establish a standard curve: the cu rve equation will display on the screen simultaneously,you can test your unknown concentration solutions by the curve.
2.2 Coefficents Methods
If you have known the coefficients K&b of the formula C=K* A+b,you can input the value directly by button,then test your unknown solution.
Specifications:
Model | V-1200 |
Wavelength Range | 325-1000nm |
Spectral Bandwidth | 4nm |
Optical System | Single Beam,Grating 1200 lines/mm |
Wavelength Accuracy | ±2nm |
Wavelength Repeatability | 0.8nm |
Photometric Accuray | ±0.5%T or ±0.004A@1A |
Photometric Range | 0-200%T,-0.3-3A,0-9999Conc |
Stray Light | 0.2%T |
Stability | ±0.002A/h@500nm |
Detector | Silicon Photodiode |
Standard Cell Holder | 4-position 10mm cell changer |
Sample Compartment | Standard 10mm pathlength cuvette |
Light Source | Tungsten lamp |
Output | USB Port & Parallel Port (Printer) |
Power Requirement | AC 110/220V 50/60HZ |
Dimension(L*W*H) | 440*387*180mm |
Weight | 12kg |
UV-1200 SPECTROPHOTOMETER
UV-1200 spectrophotometer is the ideal instrument for education and QC laboratories.Using your standard sample solutions,you can get a standard curve on the large LCD screen.They are widely used in colleges and enterprises for general quantitative analysis and experiments.
Features
1.Large LCD screen (128*64 Dots).
2.Can display total 50 groups of data,3 groups per screen.Can display standard curve and the curve equation.
3.System can also save the test results,total 200 groups of datas and 200 standard curves can be saved,it is convenient for check and reload.
4.Data can be restored after a sudden power cut.
5.Auto setting wavelength.
6.Tungsten lamp & Deuterium lamp can be turned on/off individually to extend lifetime.
7.Pre-aligned design makes it convenient to change lamps.
8.Large sample compartment,it can accommodate 5-100mm path length cuvettes with optional holders,a variety of optional accessories are available.
9. The free application software M.Wave Basic provides complete control of the spectrophotometer through the Built-in USB port,you can achieve the following functions:
Ⅰ.Quantitative Ⅱ.Kinetics Ⅲ.Photometric Test
Functions:
1.Basic Mode:Absorbance,Transmittance,Concentration
2.Quantitative
2.1 Standard Curve
Maximum of 9 standard samples can be used to establish a standard curve: the cu rve equation will display on the screen simultaneously,you can test your unknown concentration solutions by the curve.
2.2 Coefficents Methods
If you have known the coefficients K&b of the formula C=K* A+b,you can input the value directly by button,then test your unknown solution.
Specifications:
Model | UV-1200 |
Wavelength Range | 200-1000nm |
Spectral Bandwidth | 4nm |
Optical System | Single Beam,Grating 1200 lines/mm |
Wavelength Accuracy | ±2nm |
Wavelength Repeatability | 0.8nm |
Photometric Accuray | ±0.5%T or ±0.004A@1A |
Photometric Range | 0-200%T,-0.3-3A,0-9999Conc |
Stray Light | 0.2%T |
Stability | ±0.002A/h@500nm |
Detector | Silicon Photodiode |
Standard Cell Holder | 4-position 10mm cell changer |
Sample Compartment | Standard 10mm pathlength cuvette |
Light Source | Tungsten & Deuterium lamp |
Output | USB Port & Parallel Port (Printer) |
Power Requirement | AC 110/220V 50/60HZ |
Dimension(L*W*H) | 440*387*180mm |
Weight | 14kg |
This year Shipping shows:
Packing Details: Standard export packing and neutral packing if needed, carton box, wooden case or upon request, packing carefully.
Delivery time: Within 15 working days after the receipt of payment.
Shipping Method: Air / Sea / Express: DHL TNT FedEx UPS EMS, etc.
A: We can provide the specific products needed around the world at significantly lower costs.
CHINCAN also offers a Private Labeling Program to provide high quality lab supplies at lower prices.
If you need further information regarding our OEM or Private Label programs, please contact us freely.
A: Usually, delivery time is 10-15 days depend on the order quantity & inventory.
A: Normally all CHINCAN products are supplied with a warranty of 12 Months from the date of shipment.
A: Payment 100% T/T, Western Union, L/C, Trade Assurance, Credit Card ect. in advance ,before shippment.
Данные ресурс не является интернет-магазином, а лишь содержит ссылки на международную торговую площадку Alibaba.com